Introducing the IQS-3150 Variable Attenuator

6

IQS-3150

Main Software Features

Display Modes

Both standard and self-adjusting modules give you a choice of three
display modes: Absolute, Reference, and X+B.

Attenuation

Display Mode

Description

Absolute

(default mode)

The displayed attenuation takes into account both the absolute value

and the offset value.

Displayed att. = absolute att. + offset value

Reference

The displayed attenuation value is relative to a defined reference value

(see Selecting a Display Mode on page 30).

Displayed att. = absolute att. – reference value + offset value

X+B

The displayed attenuation is the sum of two elements:

³

X = physical attenuation introduced by the unit.

³

B = correction factor (dB). You must define the correction factor

for the test wavelength before enabling the X+B mode (refer to
Setting the B Value on page 23).

Displayed att. = absolute att. + correction factor

a

+ offset value

³

B = input power value (dBm)

Displayed att. = (absolute att. × –1) + input power

a

+ offset value

a.

The correction factor and the input power value are specific to the wavelength.