beautypg.com

Output[1..n]:reference – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 140

background image

SCPI Command Reference

134

IQS-3150

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the parameter.

The parameter represents the new

power reference value.

Example(s)

INP:WAV 1310 NM

CONT:MODE POW

OUTP:ALC:STAT OFF

OUTP:APM ABS

OUTP:OFFS 0.000 DB

OUTP:RPOW -15.000 DBM

OUTP:APM REF

OUTP:RPOW? Returns: 0.000000E+000

OUTP:REF? Returns: -1.500000E+001

OUTP:REF -10.000

OUTP:RPOW? Returns: -5.000000E+000

See Also

:INPut[1..n]:REFerence

:OUTPut[1..n]:APMode

:OUTPut[1..n]:OFFSet

:OUTPut[1..n]:REFerence?

:OUTPut[1..n]:RPOWer

:OUTPut[1..n]:REFerence