SCPI Command Reference

Variable Attenuator

129

Product-Specific Commands—Description

:OUTPut[1..n]:POWer

Description

This command sets the absolute output power to

a specific value. The valid range of values

depends on the type of instrument, the

configuration, the current wavelength, and the

input power. This value is used only when the

POWer control mode is active.

At *RST, the value that will be set depends on the

instrument you have.

Syntax

:OUTPut[1..n]:POWerD

BM]>|MAXimum|MINimum|DEFault

Parameter(s)

Power:
The program data syntax for is defined

as a element followed by an

optional element.

The allowed

element is DBM. The special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

lowest supported value.

MAXimum allows to set the instrument to the

highest supported value.

DEFault allows the instrument to select a value

for the parameter.

The parameter is a valid output power

in dBm. You can use the OUTPut[1..n]:POWer?

MAX and OUTPut[1..n]:POWer? MIN queries to

determine a valid range for the output power.