Input[1..n]:rattenuation – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 109

background image

SCPI Command Reference

Variable Attenuator

103

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the parameter.

The parameter is a valid

relative attenuation value. You can use the

INPut[1..n]:RATTenuation? MAX and

INPut[1..n]:RATTenuation? MIN queries to

determine a valid range for the attenuation.

a) In ABSolute mode, =

absolute attenuation + offset value.

b) In REFerence mode,

= absolute attenuation – reference value + offset

value.

c) In XB mode (dB), =

absolute attenuation + correction factor

(wavelength-specific) + offset value.

d) In XB mode (dBm), =

(absolute attenuation *–1) + input power

(wavelength-specific) + offset value.

Example(s)

INP:WAV 1310 NM
CONT:MODE ATT

OUTP:APM ABS

INP:OFFS 1.000 DB

INP:RATT 15.355 DB

:INPut[1..n]:RATTenuation