beautypg.com

Output[1..n]:rpower – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 144

background image

SCPI Command Reference

138

IQS-3150

Product-Specific Commands—Description

a) In ABSolute mode, =

absolute power + power offset value.

b) In Reference mode, =

absolute power – power reference value +

power offset value.

c) In X+B mode (dBm), =

absolute power + correction factor

(wavelength-specific) + power offset value.

Example(s)

INP:WAV 1310 NM

CONT:MODE POW

OUTP:APM ABS

OUTP:OFFS -10.500 DB

OUTP:RPOW -40.00 DBM

STAT:OPER:BIT8:COND? Keep sending the query

as long as the ouput power is not reached. When

output power is reached, the returned value is 0.

OUTP:RPOW? Returns -4.000000E+001

OUTP:POW? Returns -2.950000E+001

OUTP:APM REF

OUTP:RPOW? Returns 0.000000E+001

OUTP:RPOW 2.00

See Also

:CONTrol[1..n]:MODE

:INPut[1..n]:RATTenuation

:OUTPut[1..n]:APMode

:OUTPut[1..n]:OFFSet

:OUTPut[1..n]:REFerence

:OUTPut[1..n]:RPOWer?

:STATus:OPERation:BIT:CONDition?

:OUTPut[1..n]:RPOWer