beautypg.com

Installing the exfo universal interface (eui), Nulling electrical offsets – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 54

background image

Operating the IQS-3150

48

IQS-3150

Installing the EXFO Universal Interface (EUI)

Installing the EXFO Universal Interface (EUI)

The EUI fixed baseplate is available for connectors with angled (APC) or
non-angled (UPC) polishing. A green border around the baseplate
indicates that it is for APC-type connectors.

To install an EUI connector adapter onto the EUI baseplate:
1.
Hold the EU connector adapter so the dust cap opens downwards.

2. Close the dust cap in order to hold the connector adapter more firmly.

3. Insert the connector adapter into the baseplate.

4. While pushing firmly, turn the connector adapter clockwise on the

baseplate to lock it in place.

Nulling Electrical Offsets

Temperature and humidity variations affect the performance of electronic
circuits and optical detectors, which can offset measurement results. To
compensate for this offset, the IQS-3150 is equipped with an offset nulling
function.

EXFO recommends performing a nulling of the electrical offsets whenever
environmental conditions change.

Bare metal

(or blue border)

indicates UPC

option

Green border

indicates APC

option

2

3

4