SCPI Command Reference

120

IQS-3150

Product-Specific Commands—Description

:OUTPut[1..n]:DTOlerance

Description

This command specifies the drift tolerance that

will be used for power tracking via the ALC

(Automatic Leveling Control) loop. This value is

only taken into account when the ALC loop is

active (OUTPut[1..n]:ALC[:STATe] ON). This

value is used only when the POWer control

mode is active.

At *RST, the value that will be set depends on the

instrument you have.

Syntax

:OUTPut[1..n]:DTOlerance

[DB]>|MAXimum|MINimum|DEFault

Parameter(s)

Drift:
The program data syntax for is defined

as a element followed by an

optional element.

The allowed

element is DB. The special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

lowest supported value.

MAXimum allows to set the instrument to the

highest supported value.