Readout – KERN & SOHN AET 200-4NM User Manual
Page 45

AET_/PET_/ILT_NM-BA-e-1410
45
13.2.3. Readout
The function enables the user to adjust balance operation to ambient
conditions (filter settings) or individual user needs. Menu
the following elements:
FILTER
Each measurement signal before being displayed is electronically
processed for the purpose of obtaining corrected parameters
specific for stable measurement result, i.e. ready to read.
Balance user can influence to some extend the range of signal processing
by selecting an appropriate FILTER value. Available options:
o
very fast
o
fast
o
average / normal
o
slow
o
very slow
While setting the filtering level the user should consider the actual
operating conditions of a balance. In case of very good operating
conditions set the filter to average/normal or fast, in case of rough
conditions set the filter to slow or very slow.
Caution:
in case of precision balances PET series the recommended range of filter
is between very fast
÷
average/normal;
in case of analytical balances and microbalances the recommended range
of filter is between average/normal
÷
very slow.
Value release
This parameter determines displaying the stability marker for a
measurement result.
There are 3 available settings of value release parameter
o
fast
o
fast + reliable /recommended/
o
reliable
Caution:
The speed of stability marker occurrence depends on applied filter value and
applied value release
Autozero function
The function is to automatically monitor and correct zero
indication of a balance.