KERN & SOHN AET 200-4NM User Manual
Page 78

78
AET_/PET_/ILT_NM-BA-e-1410
19.6. Report from completed density determination processes
On completing any type of density determining processes, i.e. solid, liquid
or air, a report is generated. The report is saved in the database <
Density reports>. The files are named by their date and time (hour) of
density determining process.
An example of report from density determination of solid.
--------------- Density ----------------
---------------- Solid ----------------
User Admin
Balance ID 400015
Date 2011.10.07
Time 10:08:09
Standard liquid Water
Temperature 22 °C
Density of standard liquid 0.9978 g/cm
3
Weighing 1 10.526 g
Weighing 2 2.586 g
Density 1.322776 g/cm
3
-------------------------------------------------
Signature
Use of data contained in the header and footer.
Should data from header and/or footer be used, follow below procedure:
o
Press <
Print header> soft key before starting density determination
procedure,
o
Carry out density determination procedure
A report is printed automatically on completing the second
measurement (Weighing 2),
When density determination result is displayed, it is possible to print
it for numerous times by multiple pressing of <
PRINT > key,
o
Press <
Print footer > soft key after completing the density
determination procedure.
Caution
:
Remember to select data to be printer in the header, footer and report. For
details see section “PRINTOUTS”.