Copy tare – KERN & SOHN AET 200-4NM User Manual
Page 113

AET_/PET_/ILT_NM-BA-e-1410
113
On completing measurements of final mass for the samples in a series, the
user can check the results in the database of series.
To check the results, enter the database of series and select a series, then a
sample and a specific sample for which
On previewing a specific weighing, a window is opened with data and values on
the weighing process. Numbers displayed next to weighing data denote
measuring cycles.
Data on a specific series can be printed or exported to a file.
24.5. Copy tare
The option enables copying once selected tare value in a series to all samples
which do not have tare value assigned, and for which differential weighing
process (status other than Weighing B) is not completed.
Means of operation:
Press key
Which displays list of samples for which
tare value is assigned.
Press tare value that should apply to
other samples in a series (which do not
have tare applied).
The software automatically applies
selected tare value to data of these
samples.