KERN & SOHN AET 200-4NM User Manual
Page 32

32
AET_/PET_/ILT_NM-BA-e-1410
Available options:
•
METHOD
It is a means for designating an accepted quality standard.
Pressing a field displays a window with a list of minimum samples
values stored in balance’s memory, following a criterion on
determining the minimum sample.
Adding a new method is available only from the menu level
•
MODE
Lock – on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. Additionally the software disables accepting the measurement
result if it is below the minimum sample limit
Warn – on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. The user can accept a measurement that is below the
minimum sample limit, but on a printout such measurement is
preceded by an asterisk (*).
Weighing with application of
If during a weighing process a user wants to be informed whether a
measurement exceeds the minimum sample limit for a specific weighing
threshold, then the
additional setting of the weighing mode.
Procedure:
1. Press grey coloured workspace
2. The display indicates menu: Settings, Keys, Information, Printouts and
Profiles
3. Select menu
4. Press < Minimum sample> field
5. Below window is opened
6. Press