Adjustment history – KERN & SOHN AET 200-4NM User Manual
Page 38

38
AET_/PET_/ILT_NM-BA-e-1410
Balance operator can use in a GLP report below listed information and signs:
adjustment (adjustment mode)
working mode (name of a working mode)
date, time,
user,
balance type,
balance ID,
level status,
nominal mass,
current mass
difference
temperature,
blank line,
dashes,
signature,
non-standard printout.
10.9. Adjustment history
Contains data on all carried out adjustment processes. The record is carried
out automatically. Each entry on adjustment comprises basic data on
completed process. Balance menu enables displaying the list of completed
adjustment processes, and each report is printable.
Printing a report from adjustment process. Enter submenu < Adjustment >,
and:
list. On displaying details of a record, press print soft key < > on display’s
upper bar.
Hint:
If memory for records on completed adjustments is full, then the oldest record is
automatically erased.
If internal procedures of an organization require maintaining complete
documentation from all carried out adjustment processes, then the list with
records on adjustment should be periodically printed and filed.
Searching for adjustment record
Balance enables searching for a specific record from completed adjustment
processes:
– Press search icon
and insert date of adjustment process.
Exporting data on completed adjustment process
Connect a data storage device to balance’s USB port. Press
key located in the upper right corner of balance’s display. The process is fully
automatic, and on its completion, a file with extension *.tdb is saved on a
data storage device connected to the USB port. The file is editable using Excel
spreadsheet or a text editor.