Manual printout / automatic printout – KERN & SOHN AET 200-4NM User Manual
Page 170

170
AET_/PET_/ILT_NM-BA-e-1410
33.3. Manual printout / Automatic printout
Balance enables generating manual or automatic printouts.
•
Manual printout: press
key on stabilization of indication (measurement
result),
•
Automatic printout is generated automatically, in accordance to the settings
of automatic printout (see point 10.8 of this user manual).
The content of a printout depends on setting in menu
Format of mass printout:
1
2
3
4 -12
13
14 15 16 17 18
stability
marker
space character mass space
unit
CR LF
Stability marker
[space] if stable
[?] if unstable
[!] air buoyancy correction mode is enabled
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to – occurs
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point and right justification
Unit
3 characters with left justification
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – a printout generated from a balance
on pressing
key, according to settings of
N (number of
measurements)
NO
Universal variable 1 … 5
NO
Date
NO
Net
NO
Time
NO
Tare
NO
Level status
NO
Gross
NO
Client
NO
Current result
NO
Warehouse
NO
Supplementary unit
NO
Product
NO
Mass
YES
Packaging
NO
Non-standard printout
NO