Ambient conditions – KERN & SOHN AET 200-4NM User Manual
Page 139

AET_/PET_/ILT_NM-BA-e-1410
139
Example no. 2 a balance AET 200-4NM:
No. Tare value
Minimum
sample
Means of operation
1
220.0000 g 0.5000 g The minimum sample refers to all net
weighments that are weighed in a
packaging with optional mass from full
measuring range of a balance (the
button applied), as well as on weighing
samples without the packaging (the
Example no. 3 a balance AET 200-4NM:
No. Tare value
Minimum
sample
Means of operation
1
0.0000 g
0.2500 g The minimum sample refers to all net
weighments that are weighed without a
packaging (the
applied).
The software identifies this setting as the
minimum sample is valid only for weighing
samples without a packaging. If tarring
option is applied then the software disables
the pictogram informing on minimum
sample in use, and renders the weighment
as carried out with tare applied, that does
not have a minimum sample criterion
determined.
The user can preview entered data, but data editing is disabled.
27.13. Ambient conditions
Database of ambient conditions contains parameters related to measurement
of ambient conditions at a workstation. Depending on balance’s configuration,
a report may contain data on temperature, humidity and value of atmospheric
pressure. In case a THB ambient conditions module is connected to a balance,
records from the module are also saved.
Procedure
•
Enter submenu < Databases>, and press <
Ambient conditions >
field
•
Press field of a desired report record, unless it is visible on the top of the
list, wind the list forward using navigating keys
•
Report’s name consists of date and time
Caution:
It is possible to use an option for searching a report.