*1 For the SV-1A/1H when a sample cup of 2 mL is used. For the SV-10A/10H and

SV-100A/100H when a sample cup of 45 mL is used.

*2 Repetitive

measurement

with the sensor plates remaining in the sample

*3 The value after calibration using a standard viscosity fluid at a temperature range

between 20°C and 30°C with no condensation.

In a measurement that takes a long time, perform calibration using a standard viscosity
fluid or purified water periodically, as necessary.

*4 The unit switches to Pa

⋅s.

*5 The operating temperature of each standard and optional accessory is as follow

Name Temperature

used

Sample cup-capacity 45 mL (AX-SV-33), Small sample cup-capacity
10 mL / Small sample cup cover (AX-SV-34), Sample cup-capacity 2 mL
(AX-SV-58)*, Sample cup holder-For capacity 2 mL (AX-SV-56-1/2)

0 to 120°C

Glass sample cup-capacity 13 mL (AX-SV-35), Glass sample
cup-capacity 2 mL (AX-SV-59)

0 to 230°C

Water jacket (AX-SV-37)

0 to 100°C

Glass Storage Container (AX-SV-38)*

0 to 180°C

Plastic Storage Container (AX-SV-39)* 0

to

80°C

* Lid: 80°C max

Take the operating temperature shown above into considerations when using the above
items.