7-3. How to Use

1 Attach the water jacket on the table along the guides.

Make sure that the water jacket is slid furthest along

the guides until stopped.

2 When the water jacket is used, be sure to raise or

remove the protector.

3 Use much care so that the sensor plates and the

temperature sensor do not come into contact with the

inner wall of the sample. If they are in contact, a

measurement error may occur.

4 Be sure to adjust the sample surface to the center of the narrow part of the sensor plates.

SV-1A SV-10A/SV-100A

Slide furthest
until stopped

Guide

Guide

Protector

Raise or remove
the protector

Sample

surface

Sensor

protective

cover

Sample

surface

adjustment

position

Sample

surface

Surface
locator plate
(Shaded portion)

SV-10A/SV-100A

Sensor

protective

cover

Sample

surface

adjustmen

t position