Specifications – A&D Weighing SV-100H User Manual
Page 88
86
16. SPECIFICATIONS
Measurement method
SV-1A/SV-1H SV-10A/SV-10H
SV-100A/SV-100H
Measurement method
Sine-wave Vibro Viscometer using the Tuning-fork Vibration method
Vibration frequency 30 Hz
Viscosity measuring range
0.3 to 1000 mPa
⋅s
0.3 to 10000 mPa
⋅s
1 to 100 Pa
⋅s
(1000 to 100000 mPa
⋅s)
Repeatability
*
2
1% (Standard deviation)
Measurement
accuracy
*
1
Accuracy
*
3
±5% (1 to 100 mPa⋅s)
±3%
(1 to 1000 mPa
⋅s) ±5%
(1 to 10 Pa
⋅s)
(1000 to 100000 mPa
⋅s)
Range
(mPa
⋅s)
Minimum
display
(mPa
⋅s)
Minimum
display
(Pa
⋅s)
Range
(mPa
⋅s)
Minimum
display
(mPa
⋅s)
Minimum
display
(Pa
⋅s)
Range
(Pa
⋅s)
Minimum
display
(Pa
⋅s)
0.3-10
0.01 0.0001
0.3-10
0.01 0.0001 1-10
0.01
10-100
0.1 0.0001
10-100
0.1 0.0001
10-100
0.1
100-1000
1 0.001
100-1000
1 0.001
Minimum display
1000-10000
10
*
4
0.01
Unit (Viscosity)
m
mPa
⋅s, Pa⋅s, cP, P
Pa
⋅s, P
Operating temperature
10 to 40°C (50 to 104°F)
Minimum sample amount
2 mL or more
10 mL or more
Temperature display
0 to 160°C/0.1°C, (32 to 320°F/0.1°F)
*
5
0 to 20°C/32 to 68°F:
±1°C/±1.8°F
20 to 30°C/68 to 86°F:
±0.5°C/±0.9°F
30 to 100°C/86 to 212°F:
±2°C/±3.6°F
Temperature
measurement accuracy
100 to 160°C/212 to 320°F:
±4°C/±7.2°F
Display
Vacuum fluorescent display (VFD)
Connection cable length
1.5 m (between the main unit and the display unit)
Communication RS-232C
standard
Power supply
AC adapter (Confirm that the adapter type is correct for the local voltage and
power receptacle type.)
Power consumption
Approx. 14 VA (Including the AC adapter)
Sensor unit: 112 (W) x 132 (D) x 291 (H) mm/Approx. 0.8 kg
Display unit: 238 (W) x 132 (D) x 170 (H) mm/Approx. 1.3 kg
External
dimensions/mass
Stand unit: 296 (W) x 314 (D) x 536 (H) mm/Approx. 4.6 kg
Standard accessories
AC adapter (1 pc)
Connection cable (1.5 m, 1 pc)
Carrying case
Note: The values for the SV-H series are when using the AX-SV-51 (stand set, sold separately).
*
1 to
*
5: See the next page for the detailed description.