SV-1A/SV-1H output format example

Viscosity

/

Temper-

ature

Display

Output format example

Remarks

L

LLLL

mPa・s

+00000.00,mPa s,+025.67,C

Zeroes are output for below measuring
range error.

0.30 mPa・s

+00000.30,mPa s,+025.67,C

10.0 mPa・s

+00010.00,mPa s,+025.67,C

100 mPa・s

+00100.00,mPa s,+025.67,C

1.00 Pa・s

+01000.00,mPa s,+025.67,C

For 1000 mPa

⋅s or greater, the displayed unit

is Pa

⋅s, but the output unit remains mPa⋅s.

m

mPa・s

/

°C

H Pa・s

+01200.00,mPa s,+025.67,C

1200 is output for above measuring
range error.

L Pa・s

+000.0000, Pa s,+051.23,F

Zeroes are output for below measuring
range error.

0.0003 Pa・s

+000.0003, Pa s,+051.23,F

0.0100 Pa・s

+000.0100, Pa s,+051.23,F

0.100 Pa・s

+000.1000, Pa s,+051.23,F

1.00 Pa・s

+001.0000, Pa s,+051.23,F

Pa・s

/

°F

H Pa・s

+001.2000, Pa

,+051.23,F

1.2 is output for above measuring range error.

L cP

+000.0000, cP

,+025.67,C

Zeroes are output for below measuring
range error.

0.30 cP

+00000.30, cP

,+025.67,C

10.0 cP

+00010.00, cP

,+025.67,C

100

cP

+00100.00, cP

,+025.67,C

10.0 P

+01000.00, cP

,+025.67,C

For 1000 cP or greater, the displayed
unit is P, but the output unit remains cP.

C

cP

/

°C

H P

+01200.00, cP

,+025.67,C

1200 is output for above measuring
range error.

L

P

+000.0000,

P

,+051.23,F

Zeroes are output for below measuring
range error.

0.0030 P

+000.0030,

P

,+051.23,F

0.100 P

+000.1000,

P

,+051.23,F

1.00 P

+001.0000,

P

,+051.23,F

10.0 P

+010.0000,

P

,+051.23,F

P

/

°F

H P

+012.0000,

P

,+051.23,F

12 is output for above measuring range error.

: Space (ASC 20h)