beautypg.com

A&D Weighing SV-100H User Manual

Page 59

background image

57

Other Output (5-ed)

Parameter

Settings Description

0

No output

1

Outputs remarks.

2

Outputs remarks,
Device ID information
and signature.

With D.P. format selected, whether or not to add remarks,
Device ID information or signature to the measurement data
can be selected.

For examples of output format, refer to "9-4 Data Output
Format Examples".

3

Outputs ID number.

With CSV format selected, whether or not to add ID number to
the measurement data can be selected.

For examples of output format, refer to "9-4 Data Output
Format Examples".


Pause at Data Output (pU5e)

Parameter

Settings Description

0

No pause

1

Pause
(Approx. 2 seconds)

Whether or not to take a pause of two seconds each time one
line is output can be selected, when the data are output via
RS-232C.

When MODE 3 of the optional compact printer AD-8121B is
used, select "1".


Function of portable type(H-fnc)

Parameter

Settings Description

0

Off

Do not use the function of portable type

1

• Auto hold function When measuring with the portable type, this function is used.

When the change of the displayed viscosity value is within the stabilization
range for a fixed period of averaging time, the stability value is set to hold.
If you press the STOP key again, it returns to the standby mode.

Averaging time

Stability band width

Cond 0 2 second

5t-b 0 Stability band width Small

Cond 1 4 second

5t-b 1

Cond 2 20 second

5t-b 2 Stability band width Large

2

Timer

mode

function

After the fixed time, the displayed viscosity value is set to hold.
If you press the STOP key again, it returns to the standby mode.










This manual is related to the following products: