A&D Weighing SV-100H User Manual
Page 72

70
SV-100A/SV-100H output format example
Viscosity
/
Temper-
ature
Display
Output format example
Remarks
L Pa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00000.00, Pa s
Zeroes are output for
below measuring
range error.
1.00 Pa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00001.00, Pa s
10.0 Pa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00010.00, Pa s
Pa・s
/
°C
H Pa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00120.00, Pa s
120 is output for
above measuring
range error.
L P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000000.0,
P
Zeroes are output for
below measuring
range error.
10.0 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000010.0,
P
100
P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000100.0,
P
P
/
°F
H P
LAB-12,2003/03/19,12:34:56,+051.23,F,+001200.0,
P
1200 is output for
above measuring
range error.
: Space (ASC 20h)
Output format example (2) With date and time added
Function setting
(
√=Output Blank=No output)
5-td 1 Date/time
√
5-ed 0
Device
ID number
SV-1A/SV-1H/SV-10A/SV-10H output format example
Viscosity / Temperature
Display
Output format example
M
mPa
⋅s /
°C
1.23 mPa
⋅s
,2003/03/19,12:34:56,+025.67,C,+00001.23,mPa s
: Space (ASC 20h)
SV-100A/SV-100H output format example
Viscosity / Temperature
Display
Output format example
M
Pa
⋅s /
°C
1.23 Pa
⋅s
,2003/03/19,12:34:56,+025.67,C,+00001.23, Pa s
: Space (ASC 20h)
Output format example (3) To output the measured temperature and viscosity only
Function setting
(
√=Output Blank=No output)
5-td 0 Date/time
5-ed 0
Device
ID number
Outputs in the order of date, time, temperature,
temperature unit, viscosity and viscosity unit.
The output data are 46 characters long excluding the
terminator.
Outputs in the order of temperature, temperature unit,
viscosity and viscosity unit.
The output data are 28 characters long excluding the
terminator.