9-3 Description of Items (For all models)

Condition (Cond)

The stability of the viscosity measurement results can be adjusted, taking ambient conditions such
as vibration into consideration.

Parameter

Settings Description

0

Follows the viscosity changes
quickly.
(Prone to vibration)

1

2

Follows the viscosity changes
slowly.
(Stable values)

When the viscosity value is unstable due to external
vibration, set a greater parameter.

To measure while following the rapid changes in
viscosity, set a smaller parameter.

With a smaller setting, the measurement is prone to
external vibration. Consider the ambient conditions
of the installation site.


Unit Upon Power-on (Unit)

The units of viscosity and temperature displayed when the power is turned on are specified.

SV-1A/1H/10A/10H

Parameter

Settings Description

0

m

mPa

⋅s

(Millipascal second)

1

Pa

⋅s

(Pascal second)

2

c

cP

(Centipoise)

3

P

(Poise)

°C

(Celsius)

4

mPa

⋅s

(Millipascal second)

5

Pa

⋅s

(Pascal second)

6

c

cP

(Centipoise)

7

Viscos-

ity

P

(Poise)

Temper-

ature

°F

(Fahrenheit)

In the standby mode, pressing
the MODE key switches the
viscosity unit.
mPa

⋅s ⇔ Pa⋅s, cP ⇔ P

With "fnc 0" selected, units can
be switched even during
measurement. *1

∗1 While receiving the measurement data using the graphing program RsVisco, changing the

mode using the HOLD key is not available.

For the viscosity over 1000mPa

⋅s, the unit is fixed to Pa⋅s. For the viscosity over 1000cP, the unit

is fixed to P.


Decimal Point (pnt)

Parameter

Settings Description

0

• Dot

"

.

"

1

Comma "

,

"

The decimal point format for the displayed measurement data and the
decimal point code for measurement data output via RS-232C are
specified.

With "Comma" selected, the separator for CSV format and RsVisco
format will be "

;

" (semicolon).