7 Troubleshooting

50

Figure 46 Not connected

Figure 47 Connected

If "Not connected" is the status then check the USB cable. When the NucleoCounter is

turned off this status will also be the case.

If data still is not received quit NucleoView and close the NC Receiver. Then open

NucleoView and check that the NC Receiver is also started. If necessary, restart the

computer before opening NucleoView again.

The status "Connected" is present when and only when:

The NucleoCounter is turned On, and

The USB Cable is properly connected, and

The NC Receiver is working properly

In case you are certain that all cables etc. are as required you can try to use the F51

Testimage to PC function in the NucleoCounter (only valid from firmwa re version 4.00 and

higher). This F51 will simply transfer a predefined testimage to the PC and NucleoView will

recognize the image if it is transferred.