8troubleshooting – Welch Allyn CL300 Surgical IlluminatorCL - User Manual User Manual
Page 10

1-8
Welch Allyn
English
8
Troubleshooting
Symptom
Cause
Remedy
Lamp Fails to start
No power to unit
Check power supply cord and outlet
Line voltage is <90 VAC
Refer to Hospital Engineering department
Lamp operational life exceeded
Install a new CL 100 or CL 300 Replacement
lamp
Safety interlock open
Check and tighten slotted screw on lamp
fixture access door
Lamp not in the Light Source
Install a CL 100 or CL 300 Replacement Lamp
Lamp fixutre cable not properly
seated
Check to insure that connector to lamp is
properly seated and the sockets are not bent.
Power supply failure
Contact Authorized Distributor
Lamp shuts off after
operating for a few
minutes and restarts
after cooling.
Inadequate air flow for cooling,
Light Source over temperature
Allow 2” (51mm) clearance at back of unit;
insure that air intake-outflow vents are clean
and not blocked. Allow unit to cool and then
restart
Fan failure
Contact authorized Distributor
Indicator light on and
lamp shuts off
Lamp life exceeded
Install a new CL 100 or CL 300 Replacement
Lamp
Replacement Lamp
will not seat into Light
Source
Lamp is not oriented correctly
Rotate lamp 180° and re-insert
Incorrect Replacement Lamp
Use only correct Replacement Lamp for Light
Source wattage.
No Fans
No power to unit
Check power supply outlet
Power supply failure
Contact authorized Distributor
Low output power
Damaged Micro Link
®
Fiber or
fiber bundle
Check ends of Micro Link
®
Fiber or fiber
bundle for damage. Replace with new Micro
Link
®
Fiber or fiber bundle.
Fiber bundle not connected
correctly.
Re-insert fiber bundle correctly.
Lamp near end of usable life
Install a new CL 100 or CL 300 Replacement
Lamp
Fiber bundle will not
install in Light Source
Not a compatible fiber bundle
See Section 4: Operating Instructions
Use compatible bundle design
Install/Remove Multi Link™
Expansion Port
See Section 4: Operating Instructions
Multi Link™ Expansion Port
Installed
Remove Multi Link™ Expansion Port