59

Reports

Generating Reports (Data Retrieval)

Data analysis may be performed on-line during the monitoring session based on the

bedside unit display or reports.

Reports to be printed may be downloaded at the bedside unit or at the Central Display

Station (See EarlySense Central Display Station manual). To ensure the privacy of the

patient's medical data a PIN must be entered before reports can be downloaded.

The data can be downloaded as CSV file and as a complete report in Acrobat PDF

format. In case of CSV file the data can be viewed by the physician using standard

available software e.g., Microsoft Excel. The data is presented numerically; any other

desired graph can be generated by the physician, based on the data. The PDF report

includes both numerical and trend vs. time graph that can also be printed and includes

information about Heart and Respiratory rates as well as patient's movement, in/out of

bed status, alerts history and patient turns (See Figure 45 for an example).

Warning

The data acquired by the EarlySense System should be
interpreted by a healthcare practitioner only

Downloading Patient Data and Reports

1. Insert a flash drive with at least 0.5 GB of free space into the USB slot. (If the

flash drive is already inserted it must be removed and re-inserted before

commencing the data download procedure).

2. Press the Download Report button (See Figure 5) on the front panel.

3. You will be prompted to insert a PIN (

Figure

43) and click Submit. A message

will appear on the bottom of the screen informing you that downloading has

commenced (

Figure

44)