beautypg.com

Alerts table – Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual

Page 41

background image

35

Alerts Table

The system uses the following alerts:

Message

Color

Description

High/Low Heart Rate:___

Red

Heart Rate exceeds thresholds.

High/Low Resp. Rate:___

Red

Respiration Rate exceeds thresholds.

Bed Exit

Sensitivity Settings:

1 (Low) – 6 (High)

Red

Bed Exit

Patient Out of Bed

For ____ Min

Red

System detected that patient is out of bed for a time

interval exceeding a preset threshold.

Patient Signal Lost

Red

System cannot detect patient’s signal.

Check patient and sensor location.

Extremely High Motion

Yellow

Patient Movement Level has been extremely high

during the last minute.

Patient Turn is OFF

Low Patient Movement

Yellow

Patient Movement Level has been very low during

nighttime (00:00-6:00). Consider setting Patient Turn

reminder ON.

Patient Turn is OFF

Air Mattress Motion

Detected

Yellow

Air Mattress was detected on the bed. Consider setting

Patient Turn reminder ON.