Intended use, Contraindications for use, Warnings and cautions – Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual

Page 12

background image

6

Mode 2: The unit is internally powered, continuously powered, ordinary portable

equipment with applied part (IPX4).

Only equipment specified in this manual and complying with requirements of EN60601

should be connected to the system

The system complies with IEC 60601-1, IEC 60601-1-2 and IEC60601-1-4 EMC class A

device.

Intended Use

The EarlySense system is intended for continuous measurement of respiration rate, heart

rate and movement, in an automatic contact-less manner, at home, hospital or clinic

setting. The system is indicated for use in children, adolescents and adults. The operation

of the EarlySense System has been studied in children (weight ≥ 10 Kg) and adults

(weight ≤111 Kg) during sleep and resting condition.

Note

Definition of Age for Children: Children aged 2 and above

Contraindications for Use

The EarlySense System is contraindicated for use in:

Patients in whom proper positioning cannot be achieved or maintained.

Patients who do not meet the weight limits tested or specified.

Situations where a dry environment cannot be ensured.

An MR environment

An explosive atmosphere or in the presence of flammable anesthetics or gases

Warnings and Cautions

The EarlySense System should be installed and serviced only by qualified service

personnel, authorized by EarlySense Ltd.

Mounting bed-side unit to the wall should be performed while exercising out-

most caution. DO NOT place the bed-side unit over patient’s head, to avoid

safety related conditions.