Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual
Page 42

36
Turn Counter Exceeded
____ Hours
Yellow
Threshold time interval for turning the patient expired.
Check Sensor
Reconnect/Replace Sensor
or Restart
Blue
Technical alert indicating that the sensor connection is
malfunctioning.
Try to reconnect the sensor and if problem persists
replace the sensor (See Troubleshooting page 66).
No AC Power
Operating on Battery
Backup
Blue
An alert indicating that the EarlySense Bedside unit is
disconnected from the electrical power outlet.
Reconnect the EarlySense bedside unit to the
electricity Mains – otherwise the bed-side will
shutdown (See Troubleshooting page 66).
Very Low Battery
System Shutting Down
Blue
An alert indicating that the battery is low, and system
is about to shut down.
Reconnect the control unit to the electricity Mains (See
Troubleshooting page 66).
System Malfunction
Please Contact Service
Engineer
Blue
Technical alert indicating that the system is
malfunctioning. User should contact a service engineer
(See Troubleshooting page 66).
Multiple Alerts
Multiple alerts occur when more than one alert is triggered at the same time. For instance
if a blue system malfunction occurs while an elevated HR is in progress, this situation is
defined as a multiple alert.
There can only be one active alert at a time. If an additional alert occurs while an alert is
already active, the system will then determine which alert will be active according to the
following priority:
• Blue
• Red