beautypg.com

Alert triggers – Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual

Page 43

background image

37

• Yellow

A new alert will become active only if the currently active alert has a lower priority.

Alert Triggers

Default alert status is:

Alert Status

Heart Rate

ON

Respiratory Rate

ON

Movement

ON

Patient Turn counter

OFF

Bed Exit

OFF

Default alert threshold values are:

Low limit threshold

High limit threshold

Heart Rate

40 BPM

130 BPM

Respiratory Rate

8 Br/min

32 Br/min

Patient Turn counter

(default)

2 hours

Bed Exit (default)

Immediate – sensitivity 3

Alerting Parameters

Low Heart Rate

High Heart Rate

Time to alert for change

in HR

Max. 60 seconds

Max 60 seconds