Note

If the patient is allowed to get out of bed without the nurse being
notified, you can disable the Bed Exit alert by touching the bell in
the In Bed section of the Home Screen.

Configuring the Bed Exit Alert

Bed Exit alert can be configured in terms of alert volume and tone (See Setting Individual

Alert Volume and Tone, page 42) and sensitivity.

Setting Bed Exit Sensitivity

Press the Bed Exit Settings button on the lower right of the Home screen. The Bed Exit

Settings Area will open (Figure 26).

Figure 26: the Bed Exit Settings Area

The Bed Exit Alert can be set either to provide an immediate alert when the patient exits

the bed (varying in 6 levels of bed exit sensitivity levels) or to provide an alert when the

patient has been out of bed for a period ranging from 3-20 minutes.

Use the arrows to choose the desired delay or sensitivity setting. After 3 sec the arrows

will disappear and the change will be applied.

• If you want an immediate alert when a patient exits the bed, click the arrows

to choose a sensitivity ranging from 1-6, with 1 being the least sensitive and

6 being the most sensitive. Note that sensitivities 5 and 6 are highly sensitive

and might be triggered due to excessive movements of a patient while in bed.

• If you want the system to alert after a period of time that the patient is out of

bed, choose the time duration you would like between when the patient

leaves the bed, and when the alert sounds. Use the same set of arrows that

you use to set the sensitivity to select the number of minutes for the delayed

Bed Exit alert.

The Bed Exit alert is disabled by default. Enable the Bed Exit alert by pressing the bell

icon.