Ip1.l – Thermcraft XST-3-0-36-3V User Manual
Page 253

Installation and Operation Handbook
User Calibration
2416 Controller
7-7
Select High-point Calibration
This is the Calibration Status display, again.
Use
/
to select the parameter for the High-point Calibration
of Input 1, ‘
ip1.H’.
Press
Adjust High-point Calibration
This is the display for adjusting the High Calibration point of
Input 1. The lower readout is a live reading of the process value,
which changes as the input changes.
Feed the desired high-point calibration signal to the controller,
from the calibration source. If the lower readout does not show
this value, then use
/
to adjust the reading to the required
value.
Press
to return to the ‘
ip-List’ header.
To protect the calibration against unauthorised adjustment return
to Operator level and make sure that the calibration parameters are
hidden. Parameters are hidden using the ‘
Edit’ facility
described in Chapter 3.
Adj
1200
CAL
ip1.L