Bio-Rad iMark Microplate Absorbance Reader User Manual
Page 41
iMark Microplate Absorbance Reader Instruction Manual
35
Warning and Error of Automatic Mapping:
Warning screen:
If the total number of wells mapped exceeds 96 wells, a warning message in the left
figure appears on the screen. The calculation of the total number of mapping wells is as
follows:
(Blanks+Standards+CNs+CWs+CPs+COs+QCs+Calibrators+Sample)*Replicates
The plate mapping is executed normally and the exceeded part is invalidated.
Error screen:
If one or more mapping parameters are over the input range, an error message in the left
figure will appear on the screen.
On the third line of the LCD, the name of the parameter which is out of range and its
input range will appear.
Example: “BLK: 0 – 10”
In case there is more than one error, the first parameter of errors is indicated. By pressing
any key the user returns to the automatic mapping screen.
Example of automatic mapping:
An example of automatic mapping is as below.
1. Conditions of each parameter
Blank = 1
Standard = 8
Control CN = 2, CW = 0, CP = 2, CO = 0, QC = 0
Calibrator = 1
Sample = 32
Replicate number = 2
2. Plate mapping data
1 2 3 4 5 6 7 8 9 10 11 12
A
B
C
D
E
F
G
H
B
B
S1
S1
S2
S2
S3
S3
S4
S4
S5
S5
S6
S6
S7
S7
S8
S8
CN1
CN1
CN2
CN2
CP1
CP1
CP2
CP2
CL1
CL1
X1
X1
X2
X2
X3
X3
X4
X4
X5
X5
X6
X6
X7
X7
X8
X8
X9
X9
X10
X10
X11
X11
X12
X12
X13
X13
X14
X14
X15
X15
X16
X16
X17
X17
X18
X18
X19
X19
X20
X20
X21
X21
X22
X22
X23
X23
X24
X24
X25
X25
X26
X26
X27
X27
X28
X28
X29
X29
X30
X30
X31
X31
X32
X32
E
E
E
E
Warning: The total
Of mapped wells has
Exceeded 96.
Press any key
Entered value is
Out of range
BLK: 0 – 10
Press any key