Tesee screening report – Bio-Rad iMark Microplate Absorbance Reader User Manual
Page 19
iMark Microplate Absorbance Reader Instruction Manual
13
9. TeSeE Screening Report
Analysis of the TeSeE screening assay results can be done with two pre-programmed protocols: “TeSeE”
protocol is stored in Endpoint Protocol position #5 and “TeSeE eq” is stored in Endpoint Protocol position #6 as
factory default setting. Both protocols are locked.
TeSeE screening report is generated only if these protocols are selected as current active protocol. Plate is first
validated according to negative and positive controls validation criteria; samples are then compared to the
calculated cut-off.
Criteria for plate validation are set as follows:
Criteria #1:
Individual Negative Control must be < 0.150
Number of valid Negative Control must be >= 3
Criteria #2:
Individual Negative Control must be < 1.400 * Negative Control average
Number of valid Negative Control must be >= 2
Criteria #3:
Positive Control average must be >= 1.000
Cutoff is calculated as follows:
TeSeE:
Positive cutoff
=
Negative Control average + 0.210
Negative cutoff = (Negative Control average + 0.210) * 0.900
TeSeE eq:
Positive cutoff
= Negative Control average + 0.210
Negative cutoff = Negative Control average + 0.090
Samples are interpreted as follows:
“POS”: Positive when sample OD >= Positive cutoff
“NEG”: Negative when sample OD < Negative cutoff
“???”: Gray zone: NEG cutoff =< sample OD < POS cutoff