Kinetic analysis protocol – Bio-Rad iMark Microplate Absorbance Reader User Manual
Page 22
iMark Microplate Absorbance Reader Instruction Manual
16
.Func key to switch from [ABC] to [abc] to [sym]
key to move cursor
.Func key to switch from [ABC] to [abc] to [sym]
key to move cursor
.Func key to switch from [ABC] to [abc] to [sym]
key to move cursor
Day/Month/Year, Hour:Min
Filter wavelength setting for 1-8 positions, “0” for no filter
Edit
Kinetic Analysis Protocol
Protocol
Security
Filters
Date Set
Lab Name
Change password
Change user
Lock/Unlock
Mode
Report
Calc
Mapping
Assay
Kit name
Reading
Inter
Shaking
Ph.mode
Speed
Number of reads (2-30), start delay time (0-999)
Reading interval time
Shake (First/Every/disable), Speed (High, Low, Mid), Time
Single / Dual wavelength
Read Speed (Fast/Step)
CC Limit (correlation coefficient limit)
Calc.range (calculation range of reading number)
for Linear regression Report
GALT factor k input (0- 9999)
Result = (2nd read – 1st read)*k
for GALT Report
Manual mapping
Automatic mapping
.Func key to switch [N] to [F]
[N] for sample number, [F] for sample type
General Kinetic or GALT assay
assay type selection
Absorbance, KINETIC Plots, Linear regression when General Kinetic Assay selected
GALT report when GALT Kinetic assay selected
For current user mode
Administrator or Common user
Administrator has power to switch
Special Buttons
Memory
Recall
Main
Protocol
Plate Data
Report
Protocol
End point
Kinetic
Checkmark
1-64 End point protocol storage
1-2 Kinetic protocol storage
Checkmark reader test
Printout
Returns to MAIN screen
SaveSTDC
Language
Blank,CN,CW,CP,CO,QC,Calibrator,Sample,Replicate well
Save STD Curve
Rename STD Curve
Select Language
5 choices of storage number
5 choices of STD
curve
Choices of pre-installed language for LCD and report printout
Bold for menu pages
LCD adj
and key to change contrast
Italic for program pages