iMark Microplate Absorbance Reader Instruction Manual

32


Mode Setting Procedure

Mode Setting Menu

This menu screen is entered by choosing the "Mode" in the End Point Protocol Menu screen or
the Kinetic Protocol Menu screen. It is used to set the plate reading conditions. The Up and
Down
arrow keys move the cursor to the desired item in the menu list. The Enter key seals the
selection and resumes to the next setting screen.

a. Set Photo Mode

Photo Mode

In this screen, the user chooses Single or Dual wavelength measurement and assigns the
measurement and reference (for dual wavelength measurement) filters. The assignment
of the reference filter appears for dual wavelength measurement.
The Up and Down arrow keys move the cursor to the desired parameter. The Right
arrow key chooses Single/Dual wavelength measurement and selects the installed filter
on the "Filters" of Edit Menu screen. The Enter key seals the selection and returns to the
previous screen.

b. Set Shaking

Shaking

In this screen, the user selects/deselects shaking function, sets the shaking strength, and
enters the shaking time. A shaking time of 0s indicates no shaking. The maximum
shaking time is 999 seconds.
The Up and Down arrow key moves the cursor to the desired parameters. The Right
arrow key toggles between Yes/No (Select/Deselect) shaking or selects strength:
Low/Middle/High. The Numerical keys are used to enter the shaking time. The Enter
key seals the selection and returns the user to the previous screen.

c. Set Read Mode

Read Mode

In this screen, the user chooses Fast or Step reading and also chooses Normal or
Evaluation mode.

The “[ ]”indicates the current selection. The Up and Down arrow key moves the cursor
to the desired parameters. The Right arrow key chooses Fast/Step speed or chooses
Normal/Eval mode. The Enter key seals the selection and returns to the previous screen.
The followings are the description of each read mode parameter.

Reading speed

Fast reading:

6 sec/single wavelength reading, 10 sec/dual wavelength
reading

Step reading:

15 sec/single wavelength reading, 30 sec/dual wavelength
reading


Reading mode

Normal mode:

Plate reading once

Evaluation mode: Four plate readings per one of measurement are performed and

the average data of their plate readings is generated as a plate
data.

Photometric:

Dual

Measurement:

400

nm

Reference:

655

nm

Set Shake parameters
Shake:

Yes

Speed:

Mid

Time:

199

sec(0-999)

»

■Reading speed

[Fast ] Step
•Read mode
[Normal] Eval

»

■Set Photo mode

•Set shaking
•Set read mode

Photometric:

Single

Measurement:

400

nm