Editing End Point Protocol



















Upper Limit= 2.000
Lower Limit= 0.000
input range: 0–3.500
Upper > Lower

Standard menu

»

■STD information

•Curvefit

•Recall STD Curve

»

■Cutoff •Mode

•Report •Mapping
•Limit •Kit name
•STDs

Cutoff

See Plate Map setting procedure.

See Standard setting procedure.

See Mode setting procedure.

»

[Not use] •Ratio

•Constant
•Contrl
•Formula

»

■Set photo mode

•Set shaking
•Set read mode

Select mapping mode
»

■Manual mapping

•Automatic mapping

Mapping

Mode

Report

STDs

Limit

Kit name for #01
»G.A.P. TEST IgG«

[ABC]

Kit name

See Kit name setting in Editing Kinetic Protocol.

See Cutoff setting procedure

Select “Protocol” in Edit Menu screen above, and press Enter. The End point Protocol
Menu will appear on the screen, if the current active protocol type is End point mode.

End Point Protocol Menu

»[Raw] •Mtx •Cnc

•Abs •Cut •Dif

•Lim •Crv

If “Dif” selected

[

Raw] •Mtx •Cnc

•Abs •Cut [Dif]

•Lim •Crv

»row, odd-even

Seting the Report Type

Use the Up and Down Arrow key to move the cursor, and
press the Right Arrow key to select or deselect the report
type. Selected report type is marked as “[ ]”. Press Enter
to complete the selection.

Setting the Limit Value

Use the Up and Down Arrow key to move the cursor, and enter the OD value using
the numeral keys. The OD value range is 0.000 - 3.500. Press Enter to complete the
selection.

Edit Menu

»

■Protocol •Lab.name

•Security •SaveSTDC
•Filters •Printer
•Date set

This menu screen can be accessed by pressing the Edit key in the MAIN screen.