Editing end point protocol, Seting the report type, Setting the limit value – Bio-Rad iMark Microplate Absorbance Reader User Manual
Page 31

iMark Microplate Absorbance Reader Instruction Manual
25
Editing End Point Protocol
Upper Limit= 2.000
Lower Limit= 0.000
input range: 0–3.500
Upper > Lower
Standard menu
»
■STD information
•Curvefit
•Recall STD Curve
»
■Cutoff •Mode
•Report •Mapping
•Limit •Kit name
•STDs
Cutoff
See Plate Map setting procedure.
See Standard setting procedure.
See Mode setting procedure.
»
■[Not use] •Ratio
•Constant
•Contrl
•Formula
»
■Set photo mode
•Set shaking
•Set read mode
Select mapping mode
»
■Manual mapping
•Automatic mapping
Mapping
Mode
Report
STDs
Limit
Kit name for #01
»G.A.P. TEST IgG«
↑
[ABC]
Kit name
See Kit name setting in Editing Kinetic Protocol.
See Cutoff setting procedure
Select “Protocol” in Edit Menu screen above, and press Enter. The End point Protocol
Menu will appear on the screen, if the current active protocol type is End point mode.
End Point Protocol Menu
»[Raw] •Mtx •Cnc
•Abs •Cut •Dif
•Lim •Crv
If “Dif” selected
[
Raw] •Mtx •Cnc
•Abs •Cut [Dif]
•Lim •Crv
»row, odd-even
Seting the Report Type
Use the Up and Down Arrow key to move the cursor, and
press the Right Arrow key to select or deselect the report
type. Selected report type is marked as “[ ]”. Press Enter
to complete the selection.
Setting the Limit Value
Use the Up and Down Arrow key to move the cursor, and enter the OD value using
the numeral keys. The OD value range is 0.000 - 3.500. Press Enter to complete the
selection.
Edit Menu
»
■Protocol •Lab.name
•Security •SaveSTDC
•Filters •Printer
•Date set
This menu screen can be accessed by pressing the Edit key in the MAIN screen.