Appendix a: troubleshooting – Bio-Rad Gel Doc™ EZ System User Manual
Page 85

Criterion Stain Free
TM
System
71
Appendix A: Troubleshooting
Follow the recommendations below to troubleshoot problems with the Imager.
Problem
Possible Cause
Solution
Green LED on front panel is
off
AC power cord is not
connected
Connect AC power cord to the imager and an
appropriate wall socket
Power is off
Turn the power switch on
Front panel LEDs remain
flashing
Firmware is not loading
Call Bio-Rad Technical Support for help
Image is not visible on the
monitor
No sample on sample tray
Place a sample on the tray and run protocol
again
Image is not bright enough Underdeveloped gel
If initially stopped after 2 minutes, allow the gel
to develop for 2-3 more minutes
Red LED flashing on front
panel
Sample tray not inserted all
the way
Push sample tray into the imager until you feel
the magnet pull the sample tray
Door interlock broken
Please close the door
Bulb fault
Replace all the lamps according to the
instructions in Chapter 2
Communication interrupted Make sure Imager power is on the USB cable is
connected to the PC. Restart the imager
Imaging button is pressed
on front panel while
software is not yet open
Red light will stop flashing when the software is
opened. Press the imaging button again to
initiate imaging