Improper storage of Criterion gels can produce numerous artifacts. Criterion
gels should be stored flat at 4ºC.
Avoid freezing or prolonged storage above 4ºC. If you suspect your gels have
been stored improperly, DO NOT USE THEM.
Problem
Possible Cause
Solution
Samples do not migrate
into gel
Tape at the bottom of the
cassette not removed
Remove tape
Insufficient buffer in integral
buffer chamber
Fill integral buffer chamber with 60 ml
running buffer
Insufficient lower electrode
buffer
Fill both halves of the lower buffer tank with
400 ml running buffer when running two
gels
Electrical disconnection
Check electrodes and connections
Bands “smile” across gel,
band pattern curves
upward at both sides of the
gel
Excess heating of gel
Check buffer composition
Completely fill both halves of the lower
buffer tank with 400 ml running buffer when
running two gels
Do not exceed recommended running
conditions
Skewed or distorted bands,
lateral band spreading
Excess salt in samples
Remove salts from sample by dialysis or
desalting column prior to sample
preparation
Insufficient sample buffer or
wrong formulation
Check buffer composition and dilution
instructions
Vertical streaking
Overloaded samples
Dilute sample
Selectively remove predominant protein in
the sample
Sample precipitation
Centrifuge samples to remove particulates
prior to sample loading
Gels run too fast, provide
poor resolution, and gel
temperature is too high
Running buffer is too
concentrated
Check buffer composition
Artifact bands at
60-70 kD
Possible skin keratin
contamination
Clean all dishware and wear gloves while
handling and loading gel
Filter all solutions through nitrocellulose
Use 10% iodoacetamide to eliminate keratin
bands