1 Introduction

753 Suppressor Module / 8.753.1003 Instructions for Use

6

1.4

Safety notes

1.4.1

Electrical safety

While electrical safety in the handling of the 753 Suppressor Module is

assured in the context of the specifications IEC 1010-1 (protection class
I

, degree of protection IP20), the following points should be noted:

Mains connection

Setting the mains voltage, checking the mains fuse and the mains

connection must be effected in accordance with the instructions in

section 2.2.

Opening the 753 Suppressor Module

When the 753 Suppressor Module is connected to the power supply

the instrument must neither be opened nor should parts be removed

from it, otherwise there is a danger of coming into contact with com-

ponents which are live. Always disconnect the instrument from all

voltage sources before you open it and ensure that the mains cable

is disconnected from mains connection 19

19 !

Protection against static charges

Electronic components are sensitive to static charging and can be

destroyed by discharges. Before you touch any of the components

inside the 753 Suppressor Module, you should earth yourself and any

tools you are using by touching an earthed object (e.g. housing of the

instrument or a radiator) to eliminate any static charges which exist.

1.4.2

General precautionary rules

Handling of solvents

Check all pump tubings and all input and output lines periodically for

possible leaks. Follow the relevant instructions regarding the handling

of flammable and/or toxic solvents and their disposal.

Regular exchange of pump tubings

Pump tubings are consumable material and must be replaced from

time to time (see Section 4.2.4). Suitable measures must be taken so

that any leak which might occur in the pump tubing or connections

during unattended operation will cause no damage (placing the

instrument at the bottom, collection device for any liquid which may

leak out).