beautypg.com

Metrohm 753 Suppressor Module User Manual

Page 14

background image

2.3 Connection to the 690 Ion Chromatograph

753 Suppressor Module / 8.753.1003 Instructions for Use

11

2 Connect suppressor block

Remove the plastic stopper from opening 17

17 on the rear wall

of the 690 Ion Chromatograph and push cable 23

23 mounted

on the suppressor block 21

21 through this opening.

Connect cable 23

23 to connection 17

17 “Suppressor“ of the 753

Suppressor Module (see Fig. 5).

3 Connect column to injector

Connect separating column 63

63 to capillary connection 65

65

leading to the injector according to section 3.7 of the 690 In-

structions for Use.

4 Rinse column

Rinse separating column 63

63 with eluent for about 10 min

according to section 3.7 of the 690 Instructions for Use.

5 Connect column to suppressor

Screw inlet capillary 24

24 marked with "Eluent" at suppressor

connection 22

22 (see Fig. 6) to outlet end of separating column

63

63 using a 6.2744.010 compression fitting.

6 Connect suppressor to detector block

Screw outlet capillary 29

29 marked with "Detector" at suppres-

sor connection 22

22 to a 6.2620.060 coupling using a

6.2744.010 compression fitting.

Screw inlet capillary 59

59 of detector block 54

54 to the other end

of the 6.2620.060 coupling.

1

3

2

24

24

27

27

28

28

29

29

25

25

26

26

Fig. 6:

Connections at suppressor

24

24

Suppressor inlet

capillary for eluent

25

25

Suppressor inlet

capillary for H

2

SO

4

26

26

Suppressor outlet

capillary for H

2

SO

4

27

27

Suppressor outlet

capillary for H

2

O

28

28

Suppressor inlet

capillary for H

2

O

29

29

Suppressor outlet

capillary for eluent

H

2

O

H

2

SO

4

Waste

Eluent

Detector

Waste