2.4 Connection to the 732/733 IC system

753 Suppressor Module / 8.753.1003 Instructions for Use

15

2.4

Connection to the 732/733 IC system

2.4.1

Electrical connection

In order to allow remote operation of the suppressor and pump via the

732 IC Detector, connection 17

17 of the 753 Suppressor Module must be

connected to the "Remote" connection of the 732 IC Detector by means

of the 6.2143.210 cable (available as an option) according to Fig. 8.

Both switches 22 and 66 on the front of the 753 Suppressor

Module must additionally be set to "ON" to permit remote operation.

753

Fig. 8: Connection 753 Suppressor Module –

732/733 IC system

2.4.2

Connection of the suppressor block

The 1.753.0100 suppressor block must first be inserted in the 733 IC

Separation Center and connected to the 753 Suppressor Module. The

inlet and outlet capillaries mounted on the suppressor are then con-

nected up. Proceed as follows:

6.2143.210

Cable of

suppressor block