3 Operation
753 Suppressor Module / 8.753.1003 Instructions for Use
22
3.4
Operation via 732 IC Detector
The 753 Suppressor Module cannot be remotely operated via the 732
IC Detector unless the installation has been carried out properly as
described in section 2 (mains connection, connection cable 732 –
753, attaching the pump tubing, connection to the IC system).
When operating the 753 Suppressor Module together with the 732 IC
Detector and the 733 IC Separation Center both pump and the sup-
pressor on the 753 Suppressor Module can be operated and switched
automatically.
Switch instrument on/off
POWER
The 753 Suppressor Module is switched on and off using mains
switch 13
13 on the rear of the instrument (see Fig. 2):
I
Instrument switched on
0
Instrument switched off
After the instrument has been switched on the power lamp 11
lights up to show that the instrument is ready for use.
Switch on remote control of the pump
ON
OFF
REMOTE
In order to allow remote operation via 732 IC Detector the remote
control must be switched on with switch 22.
ON
Remote control switched on
OFF
Remote control switched off
Automatic pump switch-on when the 732 IC
Detector is switched on
In order to start up the pump drive of the 753 Suppressor Module
automatically when the 732 IC Detector is switched on alterations
must be made to the basic settings (Setup, see section 4.4.1 of
732/733 Manual). The remote output lead 1 must be set to 1 (on,
active, 0 V):
>SETUP/output
remote 10000000
When the 732 IC Detector is switched off the pump at the 753
Suppressor Module is also switched off.