Rotate suppressor at an event, Suppressor position display – Metrohm 753 Suppressor Module User Manual
Page 27

3 Operation
753 Suppressor Module / 8.753.1003 Instructions for Use
24
Rotate:
>PROGRAM/edit XXX.X min
remote
T
1
TTTTTT
Reset
output lead:
>PROGRAM/edit XXX.X min
remote
T
0
TTTTTT
Example
Automatically changing the
suppressor position in "inject"-
programs (for manual operation
or with autosamplers).
PROGRAM
progr.type: inject
0.0 remote
T
T
1
TTTTTT
TTTTTT
0.1 remote
T
T
0
TTTTTT
TTTTTT
Rotate suppressor at an event
In the 732 IC Detector the
program a maximum of 4 different events (see section 4.7.3 of
732/733 Manual ). At each event the suppressor can be brought
into the next position by setting remote output lead 2 to 1 (on,
active, 0 V). This lead must later be returned to the setting 0 (off,
inactive, open) before the suppressor can be switched to the next
position.
Rotate:
EVENT YY-MM-DD HH:MM:SS
remote
T
1
TTTTTT
Reset
output lead:
EVENT YY-MM-DD HH:MM:SS
remote
T
0
TTTTTT
Suppressor position display
STEP
SUPPRESSOR
The position display 77 of the 753 Suppressor Module beside
switch 88 shows whether the suppressor is in the right position:
Lamp on
Suppressor in the right position
Lamp off
Suppressor in the wrong position or blocked
(procedure see section 4.1)