5.1 Technical data

753 Suppressor Module / 8.753.1003 Instructions for Use

35

5

Appendix

5.1

Technical data

Pump

Pump type

2-channel peristaltic pump with rotational speed

of

20/min (50 Hz)

24/min (60 Hz)

Pump capacity

typically 0.5

0.6 mL/min with

6.1826.050 pump tubing

Pressure

max. 4 bar (0.4 MPa)

Pumpable liquids

Clear liquids with no solid contents

Pump tubing material

PVC (Tygon

)

Remote interface for pump

Purpose

Remote control of pump (switching on/off)

Pin assignment

330

+5V

2

5

+

-

Pin

Remote interface for suppressor

Purpose

Remote control of suppressor (rotating)

Pin assignment

Pin

1k

+5V

2

5

+

-

10k

0V

* A precondition for triggering a step is that the

current flows for at least 10 ms. In order that a

new step can be triggered no current must flow

for at least 10 ms before.

Current flows:

on

Current does not flow:

off

external

753

Current flows:

Step

*

Current does not flow:

external