Metrohm 753 Suppressor Module User Manual
Page 16

2.3 Connection to the 690 Ion Chromatograph
753 Suppressor Module / 8.753.1003 Instructions for Use
13
30
30
32
32
33
33
34
34
44
55
34
34
32
32
25/28
25/28
10
10
31
31
31
31
Fig. 7: Installing pump tubings
44
Tubing cartridge
30
30
PTFE tubing (6.1803.020)
55
Contact pressure lever
31
31
PEEK compression fitting
(6.2744.010)
10
10
Snap-action lever
32
32
Coupling (6.2744.030)
25
25
Suppressor inlet capillary
for H
2
SO
4
33
33
Pump tubing (6.1826.050)
28
28
Suppressor inlet capillary
for H
2
O
34
34
Stopper (white-yellow)
9 Suppressor connection 3: H
2
O
•
Pull the inlet capillary 28
28 marked with “H
2
O“ at suppressor
connection 22
22 (see Fig. 6) through one of the rear panel
openings 29
29 of the 690 Ion Chromatograph or through the
door gap in the right-hand lower corner of the interior door
13
13.
•
Mount a PEEK compression fitting 31
31 (6.2744.010) on the
end of inlet capillary 28
28 and screw this on to coupling 32
32
(6.2744.030) (see Fig. 7).
•
Push coupling 32
32 on to the outlet end of the second pump
tubing 33
33 (see Fig. 7).
•
Mount suction tubing: cut a piece of PTFE Tubing 30
30
(6.1803.020) to the desired length.
•
Mount a PEEK compression fitting 31
31 (6.2744.010) on one
end of PTFE tubing 30
30 and screw this on to coupling 32
32 (see
Fig. 7).
•
Push coupling 32
32 on to the inlet end of the second pump
tubing 33
33 (see Fig. 7).
•
Immerse the other end of the suction tubing in a vessel
containing rinsing solution (normally dist. H
2
O) and fix in
place.