2.3 Connection to the 690 Ion Chromatograph

753 Suppressor Module / 8.753.1003 Instructions for Use

13

30

30

32

32

33

33

34

34

44

55

34

34

32

32

25/28

25/28

10

10

31

31

31

31

Fig. 7: Installing pump tubings

44

Tubing cartridge

30

30

PTFE tubing (6.1803.020)

55

Contact pressure lever

31

31

PEEK compression fitting

(6.2744.010)

10

10

Snap-action lever

32

32

Coupling (6.2744.030)

25

25

Suppressor inlet capillary

for H

2

SO

4

33

33

Pump tubing (6.1826.050)

28

28

Suppressor inlet capillary

for H

2

O

34

34

Stopper (white-yellow)

9 Suppressor connection 3: H

2

O

Pull the inlet capillary 28

28 marked with “H

2

O“ at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph or through the

door gap in the right-hand lower corner of the interior door

13

13.

Mount a PEEK compression fitting 31

31 (6.2744.010) on the

end of inlet capillary 28

28 and screw this on to coupling 32

32

(6.2744.030) (see Fig. 7).

Push coupling 32

32 on to the outlet end of the second pump

tubing 33

33 (see Fig. 7).

Mount suction tubing: cut a piece of PTFE Tubing 30

30

(6.1803.020) to the desired length.

Mount a PEEK compression fitting 31

31 (6.2744.010) on one

end of PTFE tubing 30

30 and screw this on to coupling 32

32 (see

Fig. 7).

Push coupling 32

32 on to the inlet end of the second pump

tubing 33

33 (see Fig. 7).

Immerse the other end of the suction tubing in a vessel

containing rinsing solution (normally dist. H

2

O) and fix in

place.