2 Installation
753 Suppressor Module / 8.753.1003 Instructions for Use
16
1 Insert suppressor block
•
733.0010 IC Separation Center (1-channel system):
Place suppressor block 21
21 to the right-hand side of detector
block 77
77 (see Fig. 14 of 732/733 Manual) on the floor of the
inner chamber in such a way that suppressor connection 22
22
points in the direction of the door.
•
733.0020 IC Separation Center (two-channel system):
Place suppressor block 21
21 on the floor of the inner chamber
(either instead of a pulsation damper 67
67 or in front of both
pulsation dampers, see Fig. 14 of 732/733 Manual).
2 Connect suppressor block
•
733.0010 IC Separation Center (one-channel system):
Remove the plastic stopper from opening 31
31 on the rear
panel of IC Separation Center 733 and push cable 23
23
mounted on the suppressor block 21
21 through this opening.
•
733.0020 IC Separation Center (two-channel system):
Remove the plastic stopper from opening 41
41 or 43
43 on the
733 IC Separation Center and push cable 23
23 mounted on the
suppressor block 21
21 through the opening.
•
Connect cable 23
23 to connection 17
17 “Suppressor“ of the 753
Suppressor Module (see Fig. 8).
3 Connect column to injector
•
see section 2.8.6 of 732/733 Manual
4 Rinse column
•
see section 2.8.6 of 732/733 Manual
5 Connect column to suppressor
•
Screw inlet capillary 24
24 marked with "Eluent" at suppressor
connection 22
22 (see Fig. 6) to outlet end of separating column
72
72 using a 6.2744.010 compression fitting.
6 Fix column
•
Insert one or two column holders 80
80 (6.2027.030, 6.2027.040
or 6.2027.050) in mounting rails 79
79 and fasten separating
column in the column holder.
7 Connect suppressor to detector block
•
Screw outlet capillary 29
29 marked with "Detector" at suppres-
sor connection 22
22 to coupling 92
92 (6.2620.060) using a
6.2744.010 compression fitting.
•
Screw inlet capillary 78
78 of detector block 77
77 to the other end
of the coupling 92
92.