2.1 Setting up the instrument

753 Suppressor Module / 8.753.1003 Instructions for Use

7

2

Installation

2.1

Setting up the instrument

2.1.1

Packaging

The 753 Suppressor Module is supplied together with the separately

packed accessories in special packagings containing shock-absorbing

foam linings designed to provide excellent protection. The actual in-

strument is packed in an evacuated polyethylene bag to prevent the in-

gress of dust. Please store all these special packagings as only they

can assure damage-free transport of the instrument.

2.1.2

Check

After receipt, immediately check whether the shipment is complete and

has arrived without damage (compare with delivery note and list of

accessories in section 5.2). In the case of transport damage, see

instructions in section 5.4.1 "Warranty".

2.1.3

Location

Position the instruments in the laboratory at a location convenient for

operation, free from vibrations and protected against a corrosive

atmosphere and contamination by chemicals.

2.1.4

Arrangement of the instruments

The best position for the 753 Suppressor Module is beside the 690 Ion

Chromatograph and below the 709 IC Pump.

The 753 Suppressor Module should always be placed at the bottom

of the instrument stack so that any leaks which may occur in the

pump tubing or connections cannot cause damage to the other

instruments by leakage of corrosive liquids (e.g. H

2

SO

4

).