2 Installation

753 Suppressor Module / 8.753.1003 Instructions for Use

12

7 Attach pump tubings

Loosen both tubing cartridges 44 from the holding clamp 99 by

pressing down snap-action lever 10

10 and remove from

mounting pin 12

12 on the 753 Suppressor Module (see Fig. 1).

Press contact pressure lever 55 on both tubing cartridges

down as far as it will go.

Insert a length of pump tubing 33

33 (6.1826.050) into each of

the tubing cartridges as shown in Fig. 7. The white-yellow

stopper 34

34 must click into the corresponding holder on the

left-hand side of the tubing cartridge.

Place the tubing cartridges on mounting pin 12

12 and press

down on the right-hand side until snap-action lever 10

10 clicks

into position on holding clamp 99. Take care that no kinks are

formed in the pump tubing.

8 Suppressor connection 2: H

2

SO

4

Pull the inlet capillary 25

25 marked with “H

2

SO

4

“ at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph or through the

door gap in the right-hand lower corner of the interior door

13

13.

Mount a PEEK compression fitting 31

31 (6.2744.010) on the

end of inlet capillary 25

25 and screw this on to coupling 32

32

(6.2744.030) (see Fig. 7).

Push coupling 32

32 on to the outlet end of the first pump tubing

33

33 (see Fig. 7).

Mount suction tubing: cut a piece of PTFE Tubing 30

30

(6.1803.020) to the desired length.

Mount a PEEK compression fitting 31

31 (6.2744.010) on one

end of PTFE tubing 30

30 and screw this on to coupling 32

32 (see

Fig. 7).

Push coupling 32

32 on to the inlet end of the first pump tubing

33

33 (see Fig. 7).

Immerse the other end of the suction tubing in a vessel

containing regeneration solution (normally 20 mmol/L H

2

SO

4

)

and fix in place.

Pull outlet capillary 26

26 marked with "Waste" at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph.

Lead outlet capillary 26

26 to a sufficiently large waste container

and fix in place.