2 mains connection – Metrohm 753 Suppressor Module User Manual

Page 11

background image

2 Installation

753 Suppressor Module / 8.753.1003 Instructions for Use

8

2.2

Mains connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with the mains voltage set wrongly and/or

wrong mains fuse there is a danger of fire!

2.2.1

Setting the mains voltage

Before switching on the 753 Suppressor Module for the first time, check

that the mains voltage set on the instrument (see Fig. 2) matches the

local mains voltage. If this is not the case, you must reset the mains

voltage on the instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 19

19 of the

753 Suppressor Module.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 20

20 below the mains

connection 19

19 and take out completely.

3 Check fuse

Carefully take the fuse installed for the desired mains voltage out

of fuse holder 20

20 and check its specifications (the position of the

fuse in the fuse holder is marked by the white arrow imprinted

next to the mains voltage range):

100

…120 V 0.5 A (slow-blow)

Metrohm No. U.600.0013

220

…240 V 0.25 A (slow-blow) Metrohm No. U.600.0010

4 Insert fuse

Change fuse if necessary and reinsert in fuse holder 20

20.

5 Install fuse holder

Depending on the desired mains voltage, insert fuse holder 20

20 in

the 753 Suppressor Module so that the corresponding mains

voltage range can be read normally and the adjacent white

arrow points to the white bar imprinted below the fuse holder

(see Fig. 4).