3.3 Operation via 690 Ion Chromatograph

753 Suppressor Module / 8.753.1003 Instructions for Use

21

3.3

Operation via 690 Ion Chromatograph

The 753 Suppressor Module cannot be remotely operated via the 690

Ion Chromatograph unless the installation has been carried out prop-

erly as described in section 2 (mains connection, connection cable

690 – 753, attaching the pump tubing, connection to the IC system).

When operating the 753 Suppressor Module together with the 690 Ion

Chromatograph the pump on the suppressor module must still be

switched on and off manually. In contrast, the suppressor position can

be automatically changed by the 690 Ion Chromatograph.

Switch instrument on/off

POWER

The 753 Suppressor Module is switched on and off using mains

switch 13

13 on the rear of the instrument (see Fig. 2):

I

Instrument switched on

0

Instrument switched off

After the instrument has been switched on the power lamp 11

lights up to show that the instrument is ready for use.

Switch pump on/off

ON

OFF

PUMP

The drive of the pump is switched on and off manually with the

switch 33 :

ON

Pump switched on

OFF

Pump switched off

Switch on remote control of the suppressor

ON

OFF

REMOTE

In order to allow remote operation via 690 Ion Chromatograph the

remote control must be switched on with switch 66.

ON

Remote control switched on

OFF

Remote control switched off

Rotate suppressor

STEP

SUPPRESSOR

When the injection valve on the 690 Ion Chromatograph is

switched to the "Inject" position the suppressor is automatically

brought to the next position.
The position display 77 beside switch 88 shows whether the

suppressor is in the right position:

Lamp on

Suppressor in the right position

Lamp off

Suppressor in the wrong position or blocked

(procedure see section 4.1)