3 operation via 690 ion chromatograph – Metrohm 753 Suppressor Module User Manual
Page 24

3.3 Operation via 690 Ion Chromatograph
753 Suppressor Module / 8.753.1003 Instructions for Use
21
3.3
Operation via 690 Ion Chromatograph
The 753 Suppressor Module cannot be remotely operated via the 690
Ion Chromatograph unless the installation has been carried out prop-
erly as described in section 2 (mains connection, connection cable
690 – 753, attaching the pump tubing, connection to the IC system).
When operating the 753 Suppressor Module together with the 690 Ion
Chromatograph the pump on the suppressor module must still be
switched on and off manually. In contrast, the suppressor position can
be automatically changed by the 690 Ion Chromatograph.
Switch instrument on/off
POWER
The 753 Suppressor Module is switched on and off using mains
switch 13
13 on the rear of the instrument (see Fig. 2):
I
Instrument switched on
0
Instrument switched off
After the instrument has been switched on the power lamp 11
lights up to show that the instrument is ready for use.
Switch pump on/off
ON
OFF
PUMP
The drive of the pump is switched on and off manually with the
switch 33
ON
Pump switched on
OFF
Pump switched off
Switch on remote control of the suppressor
ON
OFF
REMOTE
In order to allow remote operation via 690 Ion Chromatograph the
remote control must be switched on with switch 66.
ON
Remote control switched on
OFF
Remote control switched off
Rotate suppressor
STEP
SUPPRESSOR
When the injection valve on the 690 Ion Chromatograph is
switched to the "Inject" position the suppressor is automatically
brought to the next position.
The position display 77 beside switch 88 shows whether the
suppressor is in the right position:
Lamp on
Suppressor in the right position
Lamp off
Suppressor in the wrong position or blocked
(procedure see section 4.1)