2 Installation

753 Suppressor Module / 8.753.1003 Instructions for Use

14

Pull outlet capillary 27

27 marked with "Waste" at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph.

Lead outlet capillary 27

27 to a sufficiently large waste container

and fix in place.

10 Startup of 753 Suppressor Module

Set switch 22 on 753 Suppressor Module to

"OFF".

Switch on 753 Suppressor Module with mains switch 13

13 and

set switch 33 to "ON".

Adjust the contact pressure for both tubing cartridges: press

contact pressure lever 55 upwards until the solutions just start

to be drawn in. Then press the contact pressure lever up-

wards until it clicks once more to obtain optimal contact pres-

sure.

Before switching the suppressor to the next position for the

first time rinse all three suppressor units with solution for

5 min.

Check all tubing from the storage vessels through the tubing

cartridges and the suppressor up to the waste containers for

leakage of liquids. If liquid escapes anywhere the corre-

sponding compression fitting must be tightened further or

changed.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever 10

10 on the right-hand

side if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).